Test Code ANHBCT Hepatitis B Core IgG and IgM (anti-HBc IgG,IgM), Serum
Additional Codes
Supplemental Test Code |
---|
7929 |
Methodology
Electrochemiluminescence Immunoassay (ECLIA)
Note: Result is a combination of IgG and IgM. To differentiate between IgG and IgM, order LAB549 Hepatitis B Core IgM, Serum.
Literature Reference:
Package insert: Roche Diagnostics, Elecsys Anti-HBc II. 2023-6, V2.0
Performing Laboratory
Barnes-Jewish Hospital Laboratory
Specimen Requirements
Specimen Type: Serum
Container/Tube: Red/Grey separator or Gold separator tube
Specimen Volume: Full tube
Additional Information: Specimen with obvious microbial contamination is not acceptable.
Alternate Acceptable Specimen: Gold-top (serum gel) tube. Spin down unopened (vacuum intact) tube within 45 minutes of draw and send 2 mL of serum refrigerated in original tube.
Specimen Transport Temperature
Ambient
Reference Values
Nonreactive (reported as reactive or nonreactive)
Interpretation depends on clinical setting.
Literature Reference:
Package insert: Roche Diagnostics, Elecsys Anti-HBc II. 2023-6, V2.0
Day(s) Test Set Up
Performed upon receipt
Turnaround Time:
STAT: not available
Routine: 2-4 hours after receipt in laboratory
Test Classification and CPT Coding
Hepatits B core antibody (HBcAb), Total
86704
Additional Information
For BJH Laboratory Use Only
Alternate Tube for Single test only:
1 mint green (lithium heparin) tube
1 lavender-top (K2EDTA) tube
1 pink-top (K2EDTA) tube
Analyte Stability:
Ambient temperature 20 to 25° C: 7 days Test immediately or refrigerate
Refrigerated, 2 to 8° C: 14 days Separated from cells
Frozen -15 to -25° C stability: 90 days.
No more than 4 freeze/thaw cycles. Repeated freeze/thaw cycles can cause analyte deterioation.
Laboratory Processing Instructions:
Completed specimens are stored refrigerated for 5 days.
LOINC Code Information
Name | LOINC Code |
Hepatitis B Core Antibodies, Total | 83100-8 |